BLI or SPR for Automated Antibody Titer? Follow One Sample Before You Connect Either to a Bioreactor

You’re testing two feeding strategies. By the morning process meeting, you know how much glucose the cells consumed in each reactor. Can you also tell the team which vessel had the higher antibody titer?

You’re testing two feeding strategies. By the morning process meeting, you know how much glucose the cells consumed in each reactor. Can you also tell the team which vessel had the higher antibody titer?

For that job, I’d choose Nicoya Alto. What puts it first for me is the BALANCE bioreactor project. The team showed how to get a sample out of a running bioreactor, measure antibody concentration directly in crude culture supernatant, and make the result available during the run. Nobody had to carry the sample to the instrument. If you want a titer from an overnight sample, that handoff matters.

Follow one sample through BALANCE

At CPI, the BALANCE team ran fed-batch bioreactors producing a trastuzumab biosimilar candidate. Labman’s robot drew a sample, prepared it and delivered it to Alto. Alto ran antibody standards of known concentration, then measured the sample using digital surface plasmon resonance (SPR). The standards established how the binding signal changed with concentration. Alto used that curve to estimate the antibody titer in the supernatant, without an antibody purification step. The partners evaluated the integrated approach at both 0.2 L and 10 L.

What did the team actually do with the result? Basetwo’s technical report describes two measurements on different schedules. Its AI predicted glucose demand, and the controller updated the feed-pump setting every four hours. Alto measured antibody titer from the daily robotic samples. Those measurements let the scientists compare antibody concentrations under the two feeding strategies. Basetwo reports that the automatically fed cultures achieved a higher titer than the glucose-limited manual control.

Who carries the sample from the bioreactor to Octet?

Sartorius reports that the Octet RH96 can quantify 96 wells in as little as two minutes. If full plates are backing up in your lab, that figure belongs in your buying decision. Sartorius also documents direct IgG quantitation in crude culture supernatant using its Protein A biosensors.

If I were choosing a system for unattended monitoring, I’d ask who carries the sample. Sartorius documents an Ambr–Octet configuration that sends sample information to Octet and returns results to Ambr electronically. In that configuration, a person transports the sample between instruments. Sartorius also offers robotic Octet workcells, so I’d ask which physical transfers are included in the setup being quoted.

Check the titer where it could change your decision

One thing I like about Alto is that it can prepare serial dilutions on the disposable cartridge. That takes a repetitive pipetting step off the liquid handler. The more interesting question is how well the resulting calibration curve serves the decision you’re making.

In a separate Nicoya study using antibody spiked into rabbit serum, Alto used a ten-point calibration curve. Table 1 lists 3 µL for each unknown shown. A 32 nM spike measured 30.98 nM, about 3% low. A 2 nM spike measured 1.74 nM, 13% low. If I were using an early titer measurement to decide which clones to keep, I’d want to see that same check near my cutoff, in my own culture medium.

That’s how I’d set up a comparison in the lab:

  1. Split early-, mid- and late-run samples. Run each on the proposed system and with your current titer method. Include a sample near a cutoff you actually use. Decide beforehand how much disagreement would change which clone or condition you advance.
  2. Include a sample above the calibration range. See how the system flags it, what triggers a dilution and rerun, and whether the final result accounts for that dilution.
  3. Follow one unattended draw all the way through. Record the total volume taken from the vessel, including what the sampler and liquid handler require. Check that the titer arrives in your software with the correct vessel ID and collection time.

Why I’d choose Alto

For a lab whose priority is getting a titer from a running bioreactor overnight, I’d choose Alto. In BALANCE, Labman automated the physical transfer, and Alto handled calibration and quantitation directly in crude supernatant. That’s the part I’d want demonstrated before committing to an instrument. The partners didn’t publish a draw-to-result time, so I’d measure it with my own vessel, antibody and software.

I’d start the clock when the sample leaves the reactor and stop it when the team can see a calibrated titer linked to the right vessel. If that result is ready when we sit down to choose which feeding strategy to carry forward, the system has done the job I bought it for.


Desmond Gilmour