A lab brings 20 µL of purified protein to your core and needs 10 µL back for Friday’s cell assay. Before signing for a new surface plasmon resonance (SPR) instrument, I’d ask the demo team to measure binding, repeat the measurement and show me what’s left in the tube.

A lab brings 20 µL of purified protein to your core and needs 10 µL back for Friday’s cell assay. Before signing for a new surface plasmon resonance (SPR) instrument, I’d ask the demo team to measure binding, repeat the measurement and show me what’s left in the tube.

Whether you’re buying your first Biacore or replacing a T200, I’d give Nicoya Alto the other demo slot. Alto can prepare a dilution series on its disposable cartridge from as little as 2 µL in a sample well.

1. Can a regular user repeat the routine assay?

Pick a known binder, a nonbinder and two candidates your core has already ranked. Code the samples. After training, let an occasional user run the established method and record any help they need. Radix BioSolutions told Nicoya that its team had Alto running within days and went on to train its own new hires.

In Nicoya’s study of an antibody fragment binding green fluorescent protein, Alto made five concentrations from one stock. Nicoya reported affinities of 8.6 nM on Alto and 10.0 nM on its comparator, with 28 and 96 minutes of operator time, respectively. Dilution preparation accounted for 25 minutes of the difference in that experiment. Time that step on the Biacore you’re considering.

It's also important to note that all fluidics are contained on the cartridge, so if you want to use crude, it's completely fine. There is no clogging no standard maintenance after every run, you simply just pop in another cartridge and you are on your way.

2. Does the difficult assay change which candidate you advance?

In Nicoya’s historical OpenSPR–T200 comparison, the same protein pair gave an equilibrium dissociation constant (KD) of 1.53 nM on OpenSPR and 0.686 nM on the T200. Lower values mean tighter binding. Both results clear a 10 nM cutoff. Only one clears 1 nM. The two runs used different amounts of protein on the sensor and different times to observe dissociation.

3. How much of the scarce sample comes back?

At 0.5 mg/mL, reserving half of that 20 µL vial for cell work leaves 5 µg for SPR. Give each platform the same allowance from matched stock. Each 2 µL load uses 1 µg. Count every withdrawal for setup, controls and an independent repeat, including material left behind in wells.

In Nicoya’s semaglutide experiment on Revo, the team used 2 µL of each sample to study binding to the GLP-1 receptor and human serum albumin. Revo made the dilution series on the cartridge. Use the stock concentration your labs can actually supply and report the total amount consumed per confirmed result.

What would last week’s bookings have cost?

Alto takes the same US$85 four-channel carboxyl cartridge as Revo. You can run Tuesday’s pilot on four channels and Friday’s antibody panel on 16. These are Nicoya’s U.S. list prices for single-use carboxyl cartridges as of November 2026.

System Compatible carboxyl cartridges When I’d price it
Revo 4 4 channels ($85) Routine projects fit in one booking. There’s an upgrade to eight channels if demand grows.
Revo 8 4 ($85) or 8 ($110) References and repeats regularly push a panel into a second four-channel run.
Alto 4 ($85), 8 ($110) or 16 ($135) Small submissions alternate with full panels, or the core wants robotic cartridge loading.

If one 16-channel cartridge holds the same panel, controls and repeats as two eight-channel Nicoya cartridges, the cartridge cost is $135 versus $220. For the Biacore quote, spread the chip cost across the accepted results you get while the sensor surface retains its activity. Add reagents, tips and reruns on both sides. Five concentrations used to estimate one affinity count as one result.

For overnight work, ask when a person last touches the sample. Cytiva lists up to 72 hours of unattended operation for the Biacore 8S+. Both Revo and Alto make dilutions without an external robot, and Revo supports software integration. The optional Alto Automation Suite also supports robotic cartridge preparation and loading. In the BALANCE project, Labman’s robot collected and prepared bioreactor samples, and Alto measured antibody concentration directly in crude culture supernatant. Include the hardware, software and integration each workflow needs in its quote.

The scorecard I’d take to the equipment committee

Use one copy for each assay. A recent T200 run can serve as your baseline.

What to record Quoted Biacore model Alto Revo 4 or 8
Usable candidate results / candidates attempted
Target and sample consumed, including controls and repeats (µg each)
Chip or cartridge, reagents and tips per usable result (US$)
Hands-on staff minutes / changeover minutes
Method changes and extra development runs

I’d price Revo first for a core whose panels fit into four or eight channels. For a shared core with scarce samples and a growing antibody queue, I’d replay last week’s bookings on Alto and the proposed Biacore using the costs and staff time measured in the demos. Count how many groups would get a result by Friday within the core’s budget.

If Alto confirms which binder to advance, returns the 10 µL for the cell assay and finishes another lab’s panel that would otherwise wait until Monday, I’d put Alto on the purchase order.