Why I’d Choose Chirascan Over JASCO for Antibody Lot Comparisons
In the far UV, the buffer can absorb so much light that you lose the part of the spectrum you care about. A 0.1 mm flow cell gives the light less buffer to pass through. That can let you measure the antibody in its formulation and keep the short-wavelength data that a longer cell would obscure.

Why I’d Choose Chirascan Over JASCO for Antibody Lot Comparisons
For an antibody lab that routinely measures both far- and near-UV circular dichroism (CD), I’d choose the Chirascan Q100. It supports the two flow-cell lengths Samsung Bioepis used to compare an eculizumab biosimilar with its reference product. And when two lots look different, Chirascan’s software lets you trace that difference back to the measured curves.
Why 0.1 mm and 10 mm both matter
| Documented configuration | Shortest flow cell listed for this configuration | Longest flow cell listed for this configuration |
|---|---|---|
| Chirascan Q100 | 0.1 mm | 10 mm |
| JASCO HTCD Plus with JFLC-515 holder | 0.2 mm | 2 mm |
In the far UV, the buffer can absorb so much light that you lose the part of the spectrum you care about. A 0.1 mm flow cell gives the light less buffer to pass through. That can let you measure the antibody in its formulation and keep the short-wavelength data that a longer cell would obscure.
In the near UV, the challenge is often a weak signal from the folded antibody. A 10 mm flow cell gives the light a longer path through the protein solution, strengthening the CD signal at the same antibody concentration. That gives you a way to see more detail without first concentrating the protein.
Samsung Bioepis used exactly those two pathlengths on a Q100 in its eculizumab biosimilar study. The team used 0.1 mm to compare secondary structure and 10 mm to compare features of the folded antibodies, recording triplicate scans. The National Cancer Institute’s biophysics facility also lists both lengths for its Q100 autosampler and recommends them for those measurements.
Why optical calibration matters when you compare peak heights
Two lots can have peaks in the same places and still differ in the relative heights of those peaks. When that’s the difference you’re investigating, you want the instrument to report the CD signal accurately across the spectrum.
Applied Photophysics engineers use optical calibration at multiple wavelengths during manufacture or service. The Q100 applies a wavelength-specific correction to the CD signal. That accounts for variation in the instrument’s response across the scan and gives you a sound basis for comparing band heights. Because the reference is optical, the accuracy of that calibration step doesn’t depend on someone preparing a solution at exactly the right concentration. JASCO’s published J-Series CD-scale procedure uses a chemical reference solution.
Why the detector matters when the antibody signal is weak
The Chirascan Q100 and V100 use a large-area avalanche photodiode. The standard detector on the JASCO J-1500 is a photomultiplier tube. Applied Photophysics reports that its photodiode converts incoming UV light into an electrical signal more efficiently than a conventional photomultiplier. At the same light level and scan time, that can make a weak feature easier to distinguish from noise. That matters in near-UV CD, where a small change in how an antibody folds can produce a faint change in the curve.
Applied Photophysics recorded an unsmoothed near-UV spectrum of lysozyme at 0.05 mg/mL on the Q100, using a 10 mm cell and ten minutes each for the baseline and sample. For antibody lots, the benefit of a cleaner signal is being able to tell whether a small shift between lots stands out above the variation among replicates.
Three controls help you decide which differences to trust
In a Novartis study on the Chirascan V100, two structurally similar antibodies showed a trough near 215 nm. Their clearest difference appeared at 195 nm. By mixing the antibodies in known proportions, the team calculated a 14% detection limit and a 49% quantification limit for the second antibody in that mixture. Your own antibody pair, formulation and method will determine how small a difference you can detect.
I’d use three controls when comparing lots.
- A formulation blank. Check the buffer background and sample absorbance to see which wavelengths you can trust.
- Independent preparations of one reference lot. See how much two curves can differ simply because someone diluted and loaded the sample again.
- A blinded sample with a known change. Choose the smallest change that would trigger follow-up. Record whether you can distinguish the sample from the reference replicates before revealing its identity.
For labs doing this regularly, Applied Photophysics reports 48 buffer–sample pairs in 24 hours for isothermal Q100 measurements. Plan a separate series for each flow cell and count your blanks and replicates when translating that figure into lots per day.
Analysis Workbench makes it easier to explain a flagged lot
Say one antibody lot comes back with a lower estimate of beta-sheet content. Before deciding what to do with it, you need to know how much that result depends on the way the curve was fitted. Chirascan now puts the raw spectrum, the structural estimate and the evidence for that estimate in one place.
With Analysis Workbench, you can subtract the formulation blank, average the replicates and fit the same far-UV spectrum with three published methods in one application. Overlay each fit on the measurement and you can see which features it reproduces. The residual plot shows exactly where the two curves disagree. If only one method suggests a loss of beta-sheet and its fit misses the band in question, you know where to focus before flagging the lot.
Workbench saves each processing step and its results in the same document. A second analyst can reopen it, trace the estimate back to the measured curve and see why the team made the call. Nobody has to piece the analysis together from exported spectra, screenshots and a spreadsheet.
What to keep an eye on
Nicoya plans to add batch processing, integrated higher-order structure analysis and an API so Workbench can connect with other software. The announcement doesn’t give a release date yet, so keep an eye on this one.
Imagine testing a protein against 30 compounds at three concentrations. That’s 90 conditions before you count the controls. Batch processing could apply the same analysis to each condition. Depending on what the API supports, your lab could send reviewed results to an electronic lab notebook or feed spectra, compound identities and replicate quality into an AI model to rank the next experiments. Your team could decide which steps to automate and where a scientist reviews the curve.
I’d choose the Q100 for an antibody comparability program. Its flow cells let you preserve far-UV detail in an absorbing buffer and strengthen weak near-UV signals. Workbench gives you the evidence behind a structural estimate when a lot looks different. I’d give the software roadmap real weight in that decision, too. If your lab grows from comparing two lots to screening dozens of compounds, planned batch analysis and an API could let you apply the method you already trust at that scale, with your team choosing which decisions still need a scientist.